General problems for Patient Monitor       The module problems
Malfunction instructions Possible cause Means to dispose
ECG is closed 1.This status is set "off" in menu
2.Malfunction with ECG module
3.The communicational signal line or ECG signal line connected the ECG module and connect board is loose contact
1.Set "on" in menu
2.Replace ECG module
3.Re-plug or replace communicational signal lines or ECG signal lines
SPO2 is closed 1.This status is set "off" in menu
2.Malfunction with SPO2 module
3.The communicational signal line or SPO2 signal line connected the SPO2 module and connect board is loose contact
1.Set "on" in menu
2.Replace SPO2 module
3.Re-plug or replace communicational signal lines or SPO2 signal lines
RESP is closed 1.This status is set "off" in menu
2.Malfunction with ECG module
1.Set "on" when the ECG is on.
2.Replace ECG module

NEXT>>